logo
Quality Rs485 Ce Ultrahigh Accuracy Ionization Vacuum Gauge factory
Quality Rs485 Ce Ultrahigh Accuracy Ionization Vacuum Gauge factory

Rs485 Ce Ultrahigh Accuracy Ionization Vacuum Gauge

Rs485 Ce Ultrahigh Accuracy Ionization Vacuum Gauge
Basic Properties
Country Of Origin
China
Brand Name
KYKY
Certificate
CE
PRODUCT MODEL
ZDF-11B5
Trading Properties
MOQ
1
Payment Method
T/T
Product Summary

Compound Vacuum Gauge , ZDF-62B5 , 10E5 to 10E-7 Pa , 6 loops , Rs485 Model ZDF-62B5 compound vacuum gauge is composed of 2 sets of low vacuum measuring units and 1 set of high vacuum measuring units. One unit of resistance gauge (RG 2)is used to achieve the measurement (1.0×10E5Pa~1.0×10E-7Pa...

Product Details
Highlight:

Ultrahigh Accuracy Ionization Vacuum Gauge

,

Ce Ionization Vacuum Gauge

,

Rs485 Ionization Vacuum Gauge

Type: ZJ-14 Ionization Gauge, ZJ-52T Resistance Gauge
Measure: 1E+5 To 1*E-7
Interface: Rs485, 0-5V
Accuracy: ±1%
Product Description

Compound Vacuum Gauge , ZDF-62B5 , 10E5 to 10E-7 Pa , 6 loops , Rs485

Model ZDF-62B5 compound vacuum gauge is composed of 2 sets of low vacuum measuring units and 1 set of high vacuum

measuring units. One unit of resistance gauge (RG 2)is used to achieve the measurement (1.0×10E5Pa~1.0×10E-7Pa)independently,

the unit of resistance gauge( RG1) combined with ionization gauge, as compound unit, is to achieve the continuous control and

measurement (1×10E5~1×10E-7Pa).

 

Advantages:

1. Manual or automatic mode.

2. Convenient and flexible control

3. Long up-time, stability, reliability

4. Strong anti-jam capability ( preferred for R&D and industrial production)

 

Specifications:

Compound Vacuum Gauge ZDF-62B5
Measured loop 3(resistance gauge: 2 , ionization gauge:1)
Type of Gauge ZJ-52T resistance gauge 2 sets
ZJ-14 ionization gauge 1 set
Measured range 1.0×105Pa~ 1.0×10-7 Pa
Controlled range 1.0×105Pa~ 1.0×10-7Pa
Controlled loop 6
Control mode: On or off relay control, Control point load: AC220V/3A DC28V/10A (Non-inductive load), Jog or regional control/with power off memory
Accuracy ±1%
Power supply AC 220V±10% 50Hz
Sampling time 1s
Power consumption 45W
Interface Rs485 ( Modbus-RTU)
External control of Ionization gauge PLC
Display Five-digits LED display
e.g.: 1.2E-1 refers to 1.2×10-1Pa
Dimension(W*H*D) 480*88*280 mm
Weight 4Kg